Maintenance; Periodic Maintenance; Maintenance Schedule - KLS Martin group Laser Surgical Unit diomax 1550 Instructions For Use Manual

Table of Contents

Advertisement

Instructions for Use
Laser Surgical Unit diomax
7

Maintenance

7.1

Periodic Maintenance

NOTICE
Repairs may only be performed by authorized Gebrüder Martin service technicians.
7.1.1

Maintenance Schedule

Inspection & Maintenance
Checking the EMERGENCY-STOP switch for
proper functioning
Inspection of accessory equipment
Inspection of the laser warning lamp
Use new sterile articles
Inspection of the laser fiber
Disposal of the used laser fibers
Cleaning and disinfection of the outer device
surfaces
Checking the electrical connections and
cables for potential defects
Safety Check (SC)
see section 12, page 68
Revision 2
®
1550
Maintenance interval
prior to the first operation on
each day of use
before every surgery
before every surgery
during every surgery
during every surgery
after every surgery
after every surgery
every six months
required at least once a year
To be carried out by
Hospital staff
Hospital staff
Hospital staff
Hospital staff
Hospital staff
Hospital staff
Hospital staff
Hospital technician
Martin Service Center
61