Maintenance; Periodic Maintenance - KLS Martin group LIMAX 120 Instructions For Use Manual

Surgical laser unit
Hide thumbs Also See for LIMAX 120:
Table of Contents

Advertisement

8

Maintenance

8.1

Periodic Maintenance

Repairs may only be carried out by service technicians specially authorized by
Gebrüder Martin.
Inspection & Maintenance
Checking the Emergency Stop
switch for proper functioning
Checking the accessories
Checking the laser warning lamp(s)
for proper functioning
Checking the pilot beam for coaxial
exit from focusing handpiece or
laser fiber
Checking the purge air flow
Checking the remaining life of the
VAC's main filter
Use new sterile items
Checking the laser fiber for potential
defects
Disposal of used laser fibers
Disposal of used sterile AIR filters
Disposal of used AIR tube (25 cm)
Disposal of full main filter of VAC
Cleaning and disinfecting the unit's
external surfaces
Checking the electrical connections
and cables for potential defects
Checking the cooling water level;
refill if necessary
Replacing the deionization element
Replacing the cooling water
Safety Check (SC)
see section 8.3, page 92
90
Maintenance interval
prior to the first operation on
each day of use
prior to each operation
prior to each operation
for each operation
for each operation
for each operation
for each operation
for each operation
after each operation
after each operation
after each operation
as required
after each operation
every six months
annually recommended, every
two years necessary
every 2 years (as part of SC)
every 2 years (as part of SC)
recommended once per year
Instructions for Use
Surgical Laser Unit Limax
To Be Carried out by
hospital staff
hospital staff
hospital staff
hospital staff
hospital staff
hospital staff
hospital staff
hospital staff
hospital staff
hospital staff
hospital staff
hospital staff
hospital staff
hospital staff
Martin Service Center
Martin Service Center
Martin Service Center
Martin Service Center
®
120
Revision 5