Display - KLS Martin group Laser Surgical Unit diomax 1550 Instructions For Use Manual

Table of Contents

Advertisement

5.1.6

Display

The laser display consists of the following elements:
Name (1) of the currently set laser
program – with the note modified (5)
if laser parameters were changed
compared to the stored program.
Information line (2) for information
texts or currently pending actions.
Program parameters (3) – The
following parameters can be set at the
user's choice:
— Laser power (watts)
— Operating mode
— Pulse duration (seconds)
— Pulse pause (seconds)
— Cycle (seconds)
— Cycle pause (milliseconds)
Active field (4) – The highlighted field
(orange) shows the parameter that can
currently be modified.
Laser status display (6),
see section 5.1.8 "Laser Status
Display", page 36.
34
Laser Surgical Unit diomax
The figure on the left shows the laser during
laser power setting
(item 4, highlighted in orange).
Use the rotary knob to select and adjust the
laser power, see section 5.1.7 "Rotary Knob
with Ring Light", page 35.
— By pressing the rotary knob, the laser
power adjustment is finished.
Menu section (7)
<BACK – This command is used to exit the
current menu level.
Last Used Programs – Indicates the
4 last-used programs.
Programs – Offers menus for working with
programs.
Logfile – Menu for creating log entries.
Pilot laser – Menu for setting pilot
brightness and flash frequency.
Setup – Menu for device parameter settings.
Display fields (8) for
— Avarage power – Indicates medium laser
output (W),
— Number of laser pulses,
— Duration – Indicates total treatment time
(hours, minutes and seconds),
— Total energy – Indicates the applied laser
energy (joules).
Date and time (9)
The format of date and time depends on the
selected menu language.
English:
All other languages:
Instructions for Use
®
1550
12-19-2013 03:55 pm
19.12.2013 15:55
Revision 2