5.12.6
Times
5.12.7
Saving the Modified Program
50
Laser Surgical Unit diomax
Up to four pulse parameters can be adjusted
depending on the pulse mode:
Pulse duration (5 ms – 10 s)
•
Duration of one laser pulse
Pulse pause (5 ms – 10 s)
•
Pause between two consecutive pulses in
pulse mode
Cycle (20 ms – 40 s)
•
Total duration of one cycle consisting of
several pulses (at least 2) with pulse
pauses.
Cycle pause (5 ms – 10 s)
•
Pause between two consecutive cycles
If the current program has been modified,
modified is displayed next to the program
name.
Options for saving the modified program in the
menu Programs:
Save under
•
The modified program is saved under a new
name. The original will be stored
unchanged.
Overwrite
•
The original program will be replaced by its
modified version.
Select Programs / Save under
•
— The text editor will appear.
Instructions for Use
®
1550
Revision 2