5.7

System Check

NOTICE
The starting screen must correspond to the illustration and must be checked by the user for
errors in the display.
The version number of the software can differ from device to device.
The version number must be indicated in case of service requests.
5.8

Foot Switch

In LASER READY mode the foot switch activates the device for the emission of laser light. The
foot switch also activates accessory devices that are connected to the "Accessory Port" (see
section 4.4.6 "Accessory Port", page 28).
The laser is activated by pressing the foot switch.
Laser light is emitted.
— The laser symbol appears on the display.
— The laser signal lamp is lit yellow. If a pulsed laser is used, it is lit up in intervals
concomitant with the laser light.
40
Laser Surgical Unit diomax
The laser is switched on and performs a self
check for 10 – 30 s.
The blue starting screen shows the
KLS Martin and diomax
software version, e. g. 1.00 in the example
on the left.
After the system check the screen displays
the standard program.
The casing (2) of the foot switch (1) securely
rests on the floor and is connected to the foot
switch socket (section 4.4.4 "Foot Switch
Connection", page 25) on the rear of the device
(item 4, section 4.4 "Rear of Device", page 24)
via the connection cable (3).