Appendix
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
82
Cautions and recommended electromagnetic environment
regarding portable and mobile RF communications
equipment such as cellular phones
IEC 60601-1-2
test level
3 Vrms
3 V (V
(150 kHz – 80 MHz)
3 V/m
3 V/m (E
(80 MHz – 2.5 GHz)
EVIS EXERA TJF TYPE 160VR OPERATION MANUAL
Compliance
level
Formula for recommended separation distance
(V
=E
=3 according to the compliance level)
1
1
)
1
d
=
)
1
d
=
d
=
Where "P" is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and "d" is the recommended separation
distance in meters (m).
This instrument complies with the requirements of
IEC 60601-1-2: 2001. However, under electromagnetic
environment that exceeds its noise level, electromagnetic
interference may occur on this instrument.
Electromagnetic interference may occur on this instrument
near a high-frequency electrosurgical equipment and/or other
equipment marked with the following symbol:
Guidance
3.5
------ -
P
V
1
3.5
------ -
P
E
1
80 MHz – 800 MHz
7
----- -
P
E
1
800 MHz – 2.5 GHz