Inspection of the biopsy valve
Normal
Slit
Cap
Main body
EVIS EXERA TJF TYPE 160VR OPERATION MANUAL
The biopsy valve is a consumable item that should be
inspected before each use. Replace it with a new one if
irregularities are observed by following inspection. An
irregular, abnormal or damaged valve can reduce the efficacy
of the endoscope's suction system, and may leak or spray
patient debris or fluids, posing an infection-control risk.
1.
Confirm that the slit and hole on the biopsy valves have no splits, cracks,
deformation, discoloration or other damage (see Figure 3.8).
Hole
Figure 3.8
2.
Attach the cap to the main body (see Figure 3.9).
Cap
Figure 3.9
Splits, cracks
Discoloration
Slit
Main body
Chapter 3 Preparation and Inspection
Abnormal
Discoloration
25