5.2
Withdrawal of the endoscope with an
abnormality
When the endoscopic image appears on the monitor
EVIS EXERA TJF TYPE 160VR OPERATION MANUAL
Distal cover
Irregularity
description
The distal cover cannot
be attached.
Others
Irregularity
description
The remote switch does
not work.
If an abnormality occurs while the endoscope is in use, take a proper measure
as described in either "When the endoscopic image appears on the monitor" or
"When the endoscopic image does not appear on the monitor or the frozen
image cannot be restored" below. After withdrawal, return the endoscope for
repair as described in Section 5.3, "Returning the endoscope for repair".
If the endoscope or endo-therapy accessory cannot be
withdrawn from the patient smoothly, do not attempt to
forcibly withdraw it. If any irregularities are suspected,
immediately contact Olympus. Forcibly withdrawing the
endoscope or endo-therapy accessory may cause patient
injury, bleeding and/or perforation.
1.
Turn OFF all equipment except the video system center, light source and
monitor.
2.
When using the image magnification function of the video system center,
release the function.
Possible cause
An improper distal cover
(not MAJ-311) is used.
Possible cause
The wrong remote
switch is operated.
The remote switch
function has been set
improperly.
Chapter 5 Troubleshooting
Solution
Use proper distal cover
(MAJ-311).
Solution
Operate the correct remote
switch.
Set the remote switch function
correctly as described in the video
system center's instruction
manual.
67