Olympus EVIS EXERA TJF-160VR Instructions Manual page 39

Table of Contents

Advertisement

8.
Hold the bottom end of the distal cover and turn it to adjust the indication
mark to the straight position as shown in Figure 3.21.
Figure 3.21
9.
Confirm that the bottom end of the distal cover does not spread as shown by
the arrows in Figure 3.22, and that the white ring of the distal end is not
covered by the distal cover as shown in Figure 3.22. Stroke the distal cover
with your fingers and pay attention that the distal cover is not put over the
distal end as shown in Figure 3.23. Otherwise, the distal cover may be
damaged. When Continuing the examination with the condition like
mentioning above is confirmed, the distal cover may slip off the distal end
during the examination. Replace the distal cover with a new one.
Figure 3.22
EVIS EXERA TJF TYPE 160VR OPERATION MANUAL
Indication mark
Normal
White ring
Chapter 3 Preparation and Inspection
Abnormal
35