4 Mounting
Nozzle
14
1
> 200 mm
Fig. 5: Mounting on round vessel tops
1
Reference plane
2
Vessel center or symmetry axis
If this distance cannot be maintained, a false signal suppression
should be carried out during setup. This applies particularly if buildup
on the vessel wall is expected. In such cases, we recommend repeat-
ing the false signal suppression at a later date with existing buildup.
In vessels with conical bottom it can be advantageous to mount the
sensor in the centre of the vessel, as measurement is then possible
down to the bottom.
Fig. 6: Vessel with conical bottom
Socket pieces should be dimensioned so that the lower end of the
transducer protrudes at least 10 mm (0.394 in) out of the nozzle.
2
VEGASON 61 • Four-wire 4 ... 20 mA/HART