Operating Instructions Manual as well as the Safety Manual that comes with instruments with SIL qualification. These manuals are available in the download area of "www.vega.com". Operating instructions VEGASON 61 - Profibus PA: Document- ID 29784 Editing status of the quick setup guide: 2019-03-12...
During work on and with the device, the required personal protective equipment must always be worn. Appropriate use VEGASON 61 is a sensor for continuous level measurement. You can find detailed information about the area of application in chapter "Product description".
The environment management system is certified according to DIN EN ISO 14001. Please help us fulfil this obligation by observing the environmental instructions in this manual: • Chapter "Packaging, transport and storage" • Chapter "Disposal" VEGASON 61 • Profibus PA...
Alternatively, you can access the data via your smartphone: • Download the VEGA Tools app from the "Apple App Store" or the "Google Play Store" • Scan the Data Matrix code on the type label of the instrument or •...
Dead zone Reference plane Mounting Mount the sensor at least 200 mm (7.874 in) away from the vessel wall. > 200 mm Fig. 2: Mounting on round vessel tops Reference plane Vessel center or symmetry axis VEGASON 61 • Profibus PA...
11. Tighten the compression nut of the cable entry gland. The seal ring must completely encircle the cable 12. Screw the housing lid back on The electrical connection is finished. VEGASON 61 • Profibus PA...
I 2 C Fig. 4: Wiring plan - single chamber housing Voltage supply, signal output Wiring plan, double chamber housing Wiring plan I 2 C Fig. 5: Wiring plan - double chamber housing Voltage supply, signal output VEGASON 61 • Profibus PA...
Fig. 6: Insert display and adjustment module Note: If you intend to retrofit the instrument with a display and adjustment module for continuous measured value indication, a higher lid with an inspection glass is required. VEGASON 61 • Profibus PA...
The actual level is then calculated on the basis of these entered values. At the same time, the operating range of the sensor is limited from maximum range to the requested range. VEGASON 61 • Profibus PA...
Page 11
"Create new". Enter the actual distance from the sensor to the product surface. All false signals in this area are detected by the sensor and saved after confirming with [OK]. VEGASON 61 • Profibus PA...
0.000 m(d) 1.245 m(d) 6.789 m(d) Vessel form Linearisation curve Channel Damping Storage tank Linear PV lin. value Sensor-TAG Sensor Display Basic adjustment ▶ Display Diagnostics Service Info Displayed value Backlight ▼ PA-Out Switched off VEGASON 61 • Profibus PA...
Page 13
Copy sensor data? Enable? Info Basic adjustment Display Diagnostics Service ▶ Info Instrument type Date of manufacture Last change using PC Sensor characteristics 24. March 2015 Software version Display now? Serial number 3.80 10. April 2015 12345678 VEGASON 61 • Profibus PA...
Ʋ Non-Ex instrument 12 … 32 V DC Ʋ Ex-ia instrument 12 … 24 V DC Ʋ Ex-d-ia instrument Lighting not possible Power supply by DP/PA segment coupler Max. number of sensors non-Ex/Ex 32/10 VEGASON 61 • Profibus PA...