8 MAINTENANCE

8.1

Battery Care

Do not overcharge the battery and avoid deep discharges. To lengthen the
battery's lifetime, use the system while the battery is between 25%~75%.
The BioCon-500
avoid deep discharge, disconnect the battery from the system if it will not be used
for more than a week. When storing the battery, pre-charging to about 75% is
recommended.
8.2

Changing the Battery Modules

See
4.1 "Installing or removing battery"
8.3

Changing the Thermal Paper

See
4.2 "Changing the thermal paper"
8.4
Cleaning & Disinfection
8.4.1
Cleaning
1) Cleaning outer case (housing) of the system
a) Wipe the main body using soft cloth dampened with isopropyl alcohol or
any other appropriate hospital cleaning solution.. Do not allow liquids to
leak into the device while cleaning..
b) Thoroughly dry the device with a clean, soft cloth before re-deployment.
2) Cleaning the probe
Cleaning the surface of the Probe;
a) Disconnect the Probe from the system.
b) Wipe the probe cap using a soft cloth dampened with isopropyl
alcohol or an appropriate hospital cleaning agent.
c) Then wipe with water-dampened cloth.
d)
Dry with a soft, clean cloth.
WARNING:
TM
does draw power from the battery even while powered off. To
To avoid electric shock disconnect the system
from the AC mains and the battery.
83