4. Maintenance

Experienced service providers are available for helping you with maintenance
of the apparatus.
time. Even after the warranty period, we provide you routine or emergency
services based on the service contract or on call.
(1) Cleaning outer case (housing) of the system
- Clean the outer case of the system with a soft, little moistened cloth.
- When the case is heavily soiled by oil or dirt, use a little bit of a neutral
detergent to clean.
- Do not use chemicals (hydrochloric acid, bleach).
(2) Periodically inspect the parts that are observable to you.
(3) Always keep the environment clean and tidy to thus no dirt or foreign
substance is present around the apparatus.
(4) If the message, "BATTERY LOW RECHAGE BEFORE NEXT USE SYSTEM
WILL BE TURNED OFF" pops up even with recharging over enough time
(about 4 hours), then it indicates that the life of the battery is over. Thus, the
battery should be replaced with new one.
(5) If it seems to be an error or unreasonable value in scanned results, please
contact the manufacturer or the agent for following:
- Re-calibration for bladder scanning
- With no improvement from re-calibration, then it needs inspection and repair
from the manufacturer or the agent.