CAUTION:
8.4.2
Disinfection
1) Disinfection of the Probe
a) Clean the Probe prior to disinfection.
b) Dampen a soft cloth with disinfected solution listed in the table below.
c) Wipe the probe with a dampened cloth.
d) Air dry or towel dry with a soft, clean cloth.
e) Inspect the Probe and the cable for any damage such as cracks.
WARNING:
CAUTION:
Disinfectants Lists for the Probe disinfection
Use
any glutaraldehyde based disinfectant to disinfect the Probe. Following table
lists compatible disinfectants.
Disinfection
Solutions
Cidex
Cidex 7
Metricide 14
Metricide 28
CAUTION:
8.5

Weekly Inspection

a) Try to scan with the probe disconnected, check if the "NO SCANHEAD"
message is displayed on the main display.
b) Thoroughly inspect the probe for cracks or leakage.
Do not immerse the console or the probe.
Do not use harsh, corrosive chemicals (e.g.
hydrochloric acid, bleach).
To avoid electric shock disconnect the system
from the AC mains and the battery
Do not immerse the console or the probe.
Type
Country of Origin
Liquid
USA
Liquid
USA
Liquid
USA
Liquid
USA
Do not use Cidex Plus or Metricide Plus 30 to
disinfect the device. Cidex Plus or Metricide Plus
30 will attack and damage the plastic enclosure.
This will be considered as abuse and will void the
warranty.
84
Manufacturer
Johnson & Johnson
Johnson & Johnson
Metrex Research Inc.
Metrex Research Inc.