Charging Voltage Test/Piezo Test - Leibinger Jet 2 Manual

Table of Contents

Advertisement

Group 7

7.7.6.3 Charging voltage test/Piezo test

Figure 96
The touch field <charge test>(176) serves
the setting of the charging voltage, the field
<piezo test> (177) the setting of the
oscillator voltage.
Oscillator voltage setting: issue of a constant oscillator voltage that serves
reconciliation with the oscillation voltage booster (Poti TRM 1).
Charging voltage setting: issue of a constant charging voltage for the adjustment of
the charging voltage booster (Offset TRM 2, boost TRM3).
Release 3.00
Dangerous electrical voltage!
Measurement under voltage has carried out only by electrical
experts!
This work must only be carried out by trained personnel or by
Leibinger service technicians!
[E38]
Data Entry
Page 119
177
176
[E16]
[E39]