Installation
16
6.5

Installing the sensor modules

• The nameplate on the sensor module specifies where the sensor module is typically installed:
HP (bottom)
LP (top)
For further information see the "Function" (® ä 11)section.
• Due to the orientation of the sensor modules, there may be a shift in the zero point, i.e. when the
container is empty or partially full, the measured value does not display zero. You can correct this
zero point shift: see "Commissioning without an operating menu"(® ä 39) or the "Position
adjustment" section(® ä 44) .
• Always install the sensor module HP below the lowest measuring point.
• Always install the sensor module LP above the highest measuring point.
• Do not mount the sensor modules in the filling curtain or at a point in the tank which could be
affected by pressure pulses from an agitator.
• Do not mount the sensor modules in the suction area of a pump.
• The adjustment and functional test can be carried out more easily if you mount the sensor modules
downstream of a shutoff device.
• If a heated sensor module is cooled during the cleaning process (e.g. by cold water), a vacuum
develops for a short time, whereby moisture can penetrate the sensor through the pressure
compensation (3). If this is the case, mount the sensor with the pressure compensation (3) pointing
downwards.
• Keep the pressure compensation and GORE-TEX® filter (3) free from contamination.
• Do not clean or touch process isolating diaphragms with hard or pointed objects.
6.6

Seal for flange mounting

NOTICE
Distorted measurement results.
The seal is not allowed to press against the process isolating diaphragm as this could affect the
measurement result.
Ensure that the seal is not touching the process isolating diaphragm.