Installing The Transmitter - Endress+Hauser Proline Prosonic Flow 93 Brief Operating Instructions

Hide thumbs Also See for Proline Prosonic Flow 93:
Table of Contents

Advertisement

Proline Prosonic Flow 93
3.4

Installing the transmitter

"
Caution!
• The ambient temperature range –20 to +60 °C (–4 to +140 °F) may not be exceeded at the
mounting location. Avoid direct sunlight.
• Always install the wall-mount housing in such a way that the cable entries point downwards.
3.4.1 Direct wall mounting
35 (1.38)
2
90 (3.54)
3.4.2 Pipe mounting
Ø 20...70
(Ø 0.79...2.75)
Endress+Hauser
mm (inch)
3
1
192 (7.56)
mm (inch)
~
155 (
~ 6.1)
1.
Connection compartment
2.
Securing screws M6 (max. ø 6.5 mm (0.25"));
screw head max. ø 10.5 mm (0.4")
3.
Housing bores for securing screws
3
A0007542
"
Caution!
Danger of overheating!
If the device is mounted on a warm pipe, make sure that
the housing temperature does not exceed +60 °C
(+140 °F) which is the maximum temperature
permitted.
A0007543
Installation
9