Table of Contents

Advertisement

3 Installation

"
!
"
20
3.3
Installation
3.3.1

Installing the Promag P sensor

Caution:
x The protective covers mounted on the two sensor flanges guard the Teflon (PTFE)
lining, which is turned over the flanges. Consequently, do not remove these covers
until immediately before the sensor is installed in the pipe.
x The covers must remain in place while the device is in storage.
x Make sure that the lining is not damaged or removed from the flanges.
Note:
Bolts, nuts, seals, etc. are not included in the scope of supply and must be supplied by
the customer.
The sensor is designed for installation between the two piping flanges:
x Observe in any case the necessary screw tightening torques on Page 23 ff.
x The mounting of additional ground disks is described on Page 21.
Fig. 13:
Installing the Promag P sensor
Seals
Comply with the following instructions when installing seals:
x Measuring tube linings with PFA or PTFE o No seals are required.
x In case you use seals with DIN flanges, use only seals according to DIN 2690.
x Make sure that the seals do not protrude into the piping cross-section.
Caution:
Risk of short circuit. Do not use electrically conductive sealing compound such as gra-
phite. An electrically conductive layer could form on the inside of the measuring tube
and short-circuit the measuring signal.
Ground cable (DN 25...200)
If necessary, a special ground cable for potential equalisation can be ordered as an
accessory (see Page 59). Detailled assembly instructions o Page 36 ff.
PROline Promag 23
Endress+Hauser