Proline Prosonic Flow 93
5.1.1 Procedure
1.
Remove the cover (a) of the connection compartment.
2.
Remove the dummy cover from the cable entry (b).
3.
Route the two connecting cables (c) of channel 1 through the cable gland (d).
4.
Route the two connecting cables of channel 1 through the cable entry (b) and into the
connection compartment of the transmitter.
5.
Place the cable retaining sleeves (e) of the two connecting cables at the ground contact
terminals (f) (Detail B).
6.
Twist down the ground contact terminals (f) so that the two cable retaining sleeves (e) are
firmly seated.
7.
Screw the ground contact terminals (f) tight.
!
Note!
The Prosonic Flow P (DN 15 to 65 / ½ to 2½") is grounded via the cable gland
 ä 33.
8.
Connect the connecting cable:
– Channel 1 upstream = 1
– Channel 1 downstream = 2
– Channel 2 upstream = 3
– Channel 3 downstream = 4
9.
Spread the rubber seal (g) along the side slit with a suitable tool (e.g. a large screwdriver)
and fix both connecting cables into place.
10. Push the rubber seal (g) up into the cable entry (b).
11. Tighten the cable gland (d).
12. Fit the cover (a) on the connection compartment and screw it on.
!
Note!
The connection compartment does not have to be assembled if the transmitter is wired
(power supply and signal cable) directly afterwards.
Endress+Hauser
Wiring
31