3.3

Installation instructions

3.3.1

Installing the Promag W sensor

Note!
Bolts, nuts, seals, etc. are not included in the scope of supply and must be supplied by
the customer.
The sensor is designed for installation between the two piping flanges.
• Observe in any case the necessary screw tightening torques on Page 24 ff.
• The mounting of additional ground disks is described on Page 23.
Fig. 16:
Installing the Promag W sensor
Seals
Comply with the following instructions when installing seals:
• Hard rubber lining → additional seals are always necessary!
• Polyurethane lining → additional seals are recommended
• For DIN flanges, use only seals according to DIN 2690.
• Make sure that the seals do not protrude into the piping cross-section.
Caution!
Risk of short-circuit! Do not use electrically conductive sealing compound such as
graphite! An electrically conductive layer could form on the inside of the measuring tube
and short-circuit the measuring signal.
Ground cable (DN 25...600)
If necessary, special ground cables for potential equalisation can be ordered as
accessories for potential equalisation (see Page 64). Detailed mounting instructions →
Page 44 ff.