Endress+Hauser Proline Prosonic Flow 93 Brief Operating Instructions page 7

Hide thumbs Also See for Proline Prosonic Flow 93:
Table of Contents

Advertisement

Proline Prosonic Flow 93
A
Fig. 1: Horizontal orientation
A
Recommended orientation with upward direction of flow
B
Recommended installation range with horizontal orientation
C
Recommended installation range max. 120°
Inlet and outlet run
If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows, etc.
Compliance with the following inlet and outlet runs is recommended in order to ensure
measuring accuracy.
Fig. 2: Inlet and outlet run (top view)
1
Valve (2/3 open)
2
Pump
3
Two pipe bends in different directions
Endress+Hauser
³ 15 x DN
³ 20 x DN
1
³ 20 x DN
2
³ 15 x DN
3
C
B
³ 3 x DN
Installation
C
A0001105
A0013459
7

Advertisement

Table of Contents
loading

This manual is also suitable for:

Proline prosonic flow 93pProline prosonic flow 93w

Table of Contents