7.3

Parameter List

7.3.1

"SETUP"/"BALANCE" Menu

Parameter
Setting values
AMBIENT
V.STABLE
STABLE*
UNSTABL.
V.UNSTBL.
APP.FILT.
FINAL.RD.*
FILLING
STAB.RNG
V. ACC.
ACC.*
FAST
V. FAST
ZER./TAR.
W/O STB.
W/ STAB.*
AUTOZER.
ON*
OFF
* Factory setting
Explanation
Sets the ambient conditions to "very stable": Activates a fast change in
the weight values in the event of a load change with a high output rate.
Recommended for the following work environment:
− Very stable table near the wall
− Closed and calm room
Sets the ambient conditions to "stable". Recommended for the following work
environment:
− Stable table
− Slight movement in the room
− Slight draft
Sets the ambient conditions to "unstable": Activates the delayed change in
weight values with a reduced output rate. Recommended for the following work
environment:
− Simple office desk
− Room with moving machinery or personnel
− Slight air movement
Sets the ambient conditions to "very unstable": Activates a significantly delayed
change in the weight values and long wait for stability with a further reduction in
the output rate.
Recommended for the following work environment:
− Noticeable and slow floor vibrations
− Noticeable building vibrations
− Weighed goods moved
− Very strong air movements
Activates a filter that enables a fast change in the display for very fast load
changes. Display changes with minimal load changes (in the digit range) occur
more slowly.
Activates a filter that enables a very fast change in the display with minimal load
changes, e.g. when filling containers.
Sets the stability to "very accurate".
Sets the stability to "accurate".
Sets the stability to "fast".
Sets the stability to "very fast".
If the button is pressed: The function of the [Zero] or [Tare] button is executed
immediately.
The function of the [Zero] or [Tare] button is only executed after stability is
achieved.
Activates automatic zeroing. The display is automatically set to zero in case of
a deviation of 0 less than (X).
Deactivates automatic zeroing. Zeroing must be triggered with the [Zero]
button.
System Settings
Operating Instructions
19