Sterilisation Parameter Settings - VWR VAPOUR-Line eco 50 Instruction Manual

Automatic vertical autoclaves
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
the chamber. The temperature measured by this probe is the PRODUCT TEMPERATURE. We suggest
putting it in the coldest point in order to be sure that the sterilisation temperature has been reached. The probe
can be put inside a small bottle containing liquid that has been sterilised. In such cases, please be aware that
we recommend not to exceed 500 ml
The instrument has also a fixed thermoprobe that controls the temperature of the vessel. The temperature
measured by this probe is the VESSEL TEMPERATURE.
You can stop the sterilisation cycle at any time by pressing the STOP button.
Stops the sterilisation cycle and resets all the alarms
WARNING! STOP button is an emergency button. Pushing it will abort the cycle and will open the
valves. We don't recommend you push it unless a problem occurs during the cycle.
The same thing will happen if the instrument is turned off by the main switch.

Sterilisation parameter settings

1. Turn on the autoclave by the main switch (Pos. 1 page 7).
2. After the initial presentation the display will show Program 1 (P1)
P 1
3. To modify this programme press the ENTER button, or press the  or  buttons until the programme to be
modified is displayed and then press ENTER.
P 1
4. An arrow will be displayed next to the sterilisation temperature, press the  or  buttons until the desired
sterilisation temperature is displayed and then press ENTER.
P 1
5. An arrow will be displayed next to the sterilisation time, press the  or  buttons until the desired
sterilisation time is displayed and then press ENTER.
6. The modified data is automatically stored in the memory.
7. Press the START button to start the sterilisation cycle.
1
2
1
► 1
2
1
1
2
1
10
1 C
0 m
1 C
0 m
1 C
0 m