Setting Or Changing An Application; Running Applications (Examples); Executing The "Toggle Between Weight Units" Function; Selecting Convertible Units And Their Decimal Places - VWR avantor Premium Balance Analytic PBA Operating Instructions Manual

Table of Contents

Advertisement

Operation
8.9
Setting or Changing an
Application
Procedure
In the "APPLIC." menu, select an application,
t
e.g. "COUNT."
Press the [Confirm] button.
t
Exit the menu.
t

8.10 Running Applications (Examples)

8.10.1 Executing the "Toggle Between
Weight Units" Function
The "Toggle between weight units" function enables
a switch between a maximum of four different units. The
selected basic unit is displayed every time the device is
started (see "UNIT", Chapter "7.3.1 "SETUP"/"BALANCE"
Menu", page 19). The units can be set during the
weighing process and the decimal places can be
adjusted.
Requirements
The "Toggle between weight units" function is activated
(see Chapter "7.3.9 "APPLIC."/"WEIGH" Menu",
page 25).
The weight value must be valid.
Procedure
To switch the displayed weight unit during weighing or
t
before an application: Press the [Toggle between
weight units] button until the desired unit is displayed.
The current weight value is displayed in the selected
y
unit.
8.10.2 Selecting Convertible Units and their
Decimal Places
Procedure
Press and hold down the [Toggle between weight
t
units] button.
Select one of the parameters "Unit 1" – "Unit 4" in the
t
submenu. To do so, press the [Confirm] button.
Select the desired unit. To do so, press the [Confirm]
t
button.
Specify the decimal places for the selected unit. To do
t
so, press the [Confirm] button once again.
Select the desired number of display digits. To do so,
t
press the [Confirm] button.
30
Operating Instructions

8.10.3 Running the "COUNT" Application

The „COUNT" application is used to determine the
number of parts that have nearly the same piece weight.
As the basis for this application, the balance measures the
weight of a manually counted number of parts (reference
quantity) and determines the individual average piece
weight. The balances then uses this individual average
piece weight as a reference for determining an unknown
number of parts.
Requirements
The „COUNTING" application is selected.
Procedure
To define the reference piece count: Press the [Down]
t
button.
To define the reference piece count, press the [Up]
t
or [Down] buttons until the desired manually
counted number appears, e.g., 10.
Press the [Confirm] button.
t
Place the number of reference parts on the
t
weighing pan.
Press the [Confirm] button.
t
The average individual piece weight is now deter-
y
mined.
Remove the reference parts from the weighing pan.
t
Then place the desired sample of uncounted parts on
t
the weighing pan.
The balance calculates the actual number of parts.
y
To toggle between the display of the average piece
t
weight, the total weight and the number of parts:
Press the [Up] and [Down] buttons.
8.11
Printing Weighing Result with
ID Marking
The device can be assigned an ID number. The ID number
is exported during GLP-compliant printing.
Requirements
− The identity number is specified (see Chapter "7.3.14
"INPUT" Menu", page 26).
− The GLP-compliant printout is activated (see Chapter
"7.3.7 "DATA.OUT."/"PRNT.PAR." Menu", page 24).
Procedure
Start the printout. To do so, press the [Print] button.
t
The GLP header is printed with the ID marking of the
y
LOT ID set in the menu and the current weight value.
The [GLP] button appears in the operating display.
y
Confirm the [PRINT] button.
t
To exit the GLP printout: Press the [GLP] button.
t
The GLP footer is printed.
y