Technical Data - ProMinent Portamess 911 Conductivity Operating Manual

Table of Contents

Advertisement

Technical data

10
Technical data
Tab. 4: Measuring parameter
Measuring ranges
Tab. 5: Unit parameters
Parameter:
Display
Measuring cycle
Measuring error (± 1 digit)
Input 1 (sensor)
Input 2 (temperature)
Permitted cell constant
Cell adjustment
24
Conductivity:
Temperature:
Salinity:
TDS:
Value:
LC display 35 x 67 mm, digit height 15 mm
approx. 2 seconds
Conductivity: < 0.5% of the measured value. With conductivities of
> 500 mS/cm < 1% of the measured value.
Temperature: < 0.3 K
Multi-contact for 2-pin and 4-pin sensor with integral temperature
sensor
4 mm sockets for separate Pt 1000 / NTC (30 kQ) temperature
sensor
0.010 ... 199.9 cm–1 (adjustable)
Direct input of the cell constants.
Automatic detection of the cell constants by KCI solution 0.01 mol/l
or 0.1 mol/l,
cell adjustment with any known solutions.
0.1 µS/cm ... 1000 mS/cm
(c > 0.8 cm
)
–1
0.1 µS/cm ... 500 mS/cm
(c = 0.2 ... 0.8 cm
–1
)
0.01 µS/cm ... 199.9 µS/cm
(c < 0.2 cm
)
–1
–20.0 ... +120.0 °C / –4 ... +248 °F
nLF: 0 ... +120 °C
0.0 ... 45.0 g/kg (0 ... 30 °C)
0 ... 1999 mg/l (10 ... 40 °C)