Technical Data

14 Technical Data
Measured
variable
Free chlorine and organically bound chlorine
Area of
application
Drinking water, swimming pool water and water of a similar quality
Measuring range
CGE 2-mA-2 ppm
CGE 2-mA-10 ppm 0.1-10 mg/l
CGE 2-4P-10 ppm
pH range
5.5 - 9.5
Temperature
range
5 - 45 °C,
Temperature compensation
No sharp rises in temperature
Storage
temperature
Frost-free between 5 and 50 °C
Resolution
Corresponds with the value at the lower end of the measuring range
Max. pressure
DLG III A/B:
DGM:
Flow
DLG III and DGM in-line probe housing
Cross
sensitivity
Bromine, iodine, ozone, ClO
measurement errors
Conductivity
Test water
30 µS/cm - 10.000 µS/cm
Operating life
Membrane cap
Typically 1 year depending upon the water quality
Materials
Membrane cap:
Electrode adaptor:
Power supply
16 - 24 V DC; min 35 mA at 16 V DC
Output signal
4 - 20 mA (mA-Variant)
0 - 2 V
Enclosure rating
IP 65
32
0.02-2.0 mg/l
0.1-10 mg/l
1 bar (free flow)
3 bar (no pressure surges!)
, chloramines and other oxidising agents lead to
2
PPE
PVC
(4P-Variant)
Optimum:
Minimum:
Maximum:
Electrode holder: PMMA
Electronics: Electronic components
30 l/h
20 l/h
100 l/h
®
ProMinent