Configuration - ProMinent Portamess 911 Conductivity Operating Manual

Table of Contents

Advertisement

5

Configuration

User qualification: instructed user
n
Ä Chapter 2.2 'User qualification'
on page 8
You can change the following basic settings of
the unit in the configuration:
Cond (conductivity), SAL (salinity) or tdS
n
(Total Dissolved Solids or evaporation res‐
idue)
Calibration by input of the cell constants
n
(AutCal Off) or calibration with a calibration
solution (AutCal On)
Temperature display °C or °F
n
To activate configuration, with the unit switched
'CAL' and then press 'ON/OFF' .
off, press
'ConF' : The menu items of the configuration
menu are processed one after the other.
Pressing ⇧ and ⇩ lets you change the settings
of the respective menu item.
'CAL' takes you to the next menu item, saving
the settings displayed.
'meas' lets you exit the configuration menu at
any time. The value last displayed and possibly
changed is not saved.
Use Function to select the measuring function
[Cond] (conductivity), SAL (salinity) or tdS
(Total Dissolved Solids or evaporation residue).
Automatic or manual calibration
Select whether you wish to adjust cells by
direct input of cell constants or by calibration
with a calibration solution and automatic drift
control. Factory default setting: Direct input of
the cell constant (AutCAL OFF)
AutCAL OFF
Direct input of the cell constants (AutCAL OFF)
from 0.010 cm
–1
... 199.9 cm
setting 0.475 cm
.
–1
AutCAL ON
Automatic calibration (AutCAL On) with 0.1
molar KCl solution, 0.01 molar KCl solution or
input of the temperature-correct conductivity of
another known calibration solution.
Temperature display °C temp
The temperature can be displayed in either °C
or °F. Factory default setting: °C.
Configuration
–1
. Factory default
15