Agilent Technologies Pump Module Original User Manual page 28

Table of Contents

Advertisement

Pump Module User Guide
Installing a Weigh Station
Installing the Weigh Station
Removing the Weigh Station
26
Figure Weigh Station
00268
Bravo
weigh pad
Before you start
Determine where to install the Weigh Station. The Weigh Station must be close
enough to the Pump Module, so that the Weigh Station cable can be connected
to the Pump Module SHELF A port.
Make sure you have the following:
Sockethead screw (supplied with the Weigh Station)
M4 hex or T30 hexalobular wrench, as applicable
M5 hex wrench
WARNING
Turn off the Pump Module and the liquid-handling device or system before
you install or remove an accessory within the physical workspace.
To install a Weigh Station:
1
If you are replacing a platepad with a Weigh Station, remove the screw in
the center of the platepad with the M4 or T30 wrench, as applicable, and
then remove the platepad.
2
Place the Weigh Station at the installation location.
Make sure the side with the cable faces towards the Pump Module and that
the Weigh Station sits level.
3
Insert the supplied sockethead screw into the center of the Weigh Station
and use the M5 hex wrench to tighten the screw.
To remove a Weigh Station:
1
Disconnect the Weigh Station from the Pump Module.
2
Using the M5 hex wrench, remove the screw in the center of the Weigh
Station.
3
Remove the Weigh Station.
Pump Module User Guide