Pump Module User Guide
Autofilling options
Auto Filling Reservoir
Tip Wash Station
10
The Auto Filling Reservoir is an open tray that can be installed on a platepad
or a Weigh Station for precision liquid- level control. The reservoir can supply
reagents to single- channel pipettors and multichannel pipette heads.
Figure Auto Filling Reservoir
2
1
3
Item
Description
1
Inlet ports. Connects to the input tubing from the Pump Module that
fills the reservoir.
2
Outlet port. Connects to the output tubing from the Pump Module
that empties the reservoir.
3
Overflow drain. Uses gravity to drain excess fluid.
When properly configured in the automation software, the Pump Module
automatically fills and drains the Auto Filling Reservoir. As the figure shows,
two inlet ports are available to supply reagents to the reservoir. An outlet port
is used to pump the waste and excess fluid from the reservoir. An overflow
drain port can also be connected to drain the excess fluid. If the overflow port
is not connected, it must be plugged.
The Tip Wash Station is available in 96- chimney or 384- chimney formats.