Connecting tube B between the Pump Module and the reservoir
Pump Module User Guide
Figure Open pump head (side view) showing fully open clamp and clamp dial
00310
4
Tip Wash Station only. For the input flow (fill direction) into the Tip Wash
Station, connect the inline filter (3) to tube B between the Pump Module
and the Tip Wash Station as the preceding figure shows.
Ensure that the filter is oriented so that the three thumbscrews face
toward the Pump Module and the other side of the filter faces toward the
Tip Wash Station.
5
Insert a tubing connector (4) on each open end of tube B, as the preceding
figure shows.
To connect tube B between the Pump Module and the reservoir:
1
Cut two lengths of tube A measured for the distance between the end of
tube B and the connector on the end of tube A from the reservoir, as the
following figure shows (1).
Figure Tubing from Pump Module to reservoir
B
B
00306
2
For the fill direction:
a
Connect one length of tube A to the tube B connector or to the inline
filter (2), if applicable, on the input tubing (fill direction) from the
Pump Module.
b
Connect the other end of tube A to the quick- disconnect fitting on the
end of tube A to be connected to the reservoir input ports.
3
For the empty direction:
a
Connect the other length of tube A to the tube B connector (3) on the
empty- direction tubing going into the Pump Module.
2
B
A
1
3
B
A
Pump Module User Guide
Routing the tubing
23