Connecting tube A to the 8-Channel Wash Station
Positioning tube B in the pump heads
Pump Module User Guide
Figure Example tubing configuration for Vertical Pipetting Station with Tip Wash Station
3
Refer to the following figure for this procedure.
To connect tube A to an 8-Channel Wash Station:
1
Fill (input). Cut a length of tube A to go from the input port (1) to a
quick- disconnect fitting (2). Connect the tubing as shown in the figure.
2
Empty (output). Cut a length of tube A to go from the output port (3) to
another quick- disconnect fitting (4). Connect the tubing as shown in the
figure.
Figure Tubing configuration for the 8-Channel Wash Station
2
1
4
WARNING
Pinch hazard! Keep your fingers clear of the pump head interior when you
close the pump head cover. Keep the pump head cover closed while the pump is running.
CAUTION
Ensure that the tubing is not kinked, twisted, or stretched against the rollers.
Make sure that the tube is not crushed in the clamps.
4
1
3
Pump Module User Guide
Routing the tubing
5
2
00299
21