Precautions And Usage Limitations - NUVE NF 1200 User Manual

Multi-purpose bench top centrifuges
Table of Contents

Advertisement

ANGLE ROTOR ACCESSORIES
Rotor
Description
RS600
Angle Rotor
Set of 6 adaptors
Set of 6 adaptors
Set of 6 adaptors
Set of 6 adaptors
RS100
Angle Rotor
Set of 10
adaptors
Set of 10
adaptors
RS24
Angle Rotor
Adaptor
Adaptor
Adaptor
RS30
Angle Rotor
Adaptor
Adaptor
Adaptor

2.5 PRECAUTIONS AND USAGE LIMITATIONS

Do not use the device for any purpose other than the usage purpose.
Prior to first use, the user's manual should be read and the device is only to be used by
authorized and trained personnel. Only authorized technical personnel handle the product in
case of any failure.
The working bench should be durable to the device weight and vibration isolated.
Ensure that the rotor is placed correctly prior to usage.
According to the standard IEC 61010-2-020, anyone and any hazardous materials should not
be in the 300 mm safety zone while centrifuge is running.
Do not move the device while it is running.
Do not open the lid while rotor is spinning.
Apply the manual lid opening procedure in the case of power cut or in the case of any error.
Use only the the spare parts, rotors and accessories which are supplied by NUVE.
Load the rotor according to the explanations in the user's manual.
Start the device after ensuring the rotor is loaded correctly.
Do not use the centrifuge in areas which are in explosive danger.
Do not centrifuge the explosive, flammable, radioactive, corrosive materials and the materials
which may react with each other.
The centrifuge and the rotor are not microbiologically leak-proof. Use tubes with leak-proof
covers, if hazardous, toxic and pathogenic microorganisms are centrifuged.
Do not use corrosive materials which may be harmful for the device integrity, rotor and
accessories.
Max.Tube
Capacity
Dia(mm)
6x100ml
1x50ml
1x50 ml conical
1x30/38ml
1x15 ml conical
10x10ml
1x6ml
1x1.5/2ml
24x1.5/2ml
1x500/800µl
1x200 µl PCR
1x250/400/700
µl
30x2ml
1x500/800µl
1x200 µl PCR
1x250/400/700
µl
8
Max.Radius
(mm)
38
99
30
30.5
26
17
16
88
13
11
11
84
8
6.5
6
11
99
8
6.5
6
Max.Speed
Max
(rpm)
RCFxg
9,000
8,965
12,000
14,167
14,000
18,405
13.500
20.170

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nf 1200 rNf 800Nf 800 r

Table of Contents