Procedure; Test 5: Jitter; Test 5.1: Period Jitter; Test Specifications - HP 8114A Service Manual

Programmable pulse generator 100 v/2 a
Table of Contents

Advertisement

Test 5: Jitter

Test 5.1: Period Jitter

Test Specifications

Equipment Needed

Procedure

HP 8 I I 4A UNDER TESl
The following tests are required:
1. Period Jitter
2. Width Jitter
3. Delay Jitter
RMS-Jitter:
0.03% + 25 ps (0.05% + 25 ps for period ~100 ns)
Digitizing Oscilloscope with Accessories
Delay Line (22 ns)
Power Splitter
All cables: 50 0, coaxial, BNC, 122 cm (4 ft)
High Power Attenuator
1. Connect HP 8114A to the Scope as shown:
Equipment Set-up for Jitter Test
,
Testing the HP 8114A
3.17

Hide quick links: