HP 8114A Service Manual page 54

Programmable pulse generator 100 v/2 a
Table of Contents

Advertisement

3.18
Testing the HP 8114A
2. On the HP 8114A press TRIGGER and set up page as follows:
. 1: : 1 ; s ). j -( ; I . , 1 I _ ( / {Ii ! _ I ; ; ; F j , J ;- :i; 5 : c ;
. ; z ; : j , r " ! :; ' j , E' 7: '/ ,- , . . . , . _
following illustration:
0.00 i-Is Baselne +0.00 v
Configuring the Output Page
4. Set the Digitizing Oscilloscope HP 54121T:
Press IAUTOSCALE)
n
channel 4 to 10
Set the VOLTS/DIV of channel 4 to 10 mVldiv
n
Set OFFSET to 1V
n
Select the Timebase menu
n
Center the first positive-going
n
Delay = 29 ns)
Select the Delta V menu and turn the V markers On
n
Set the Marker 1 Position to 980 mV and the Marker 2 Position
to 1 v
the
OUTPUT page as shown in the
Out ON
Inh OFF f
and
set the TIME/DIV to 100 ps/div
edge
of the signal (approximate