HP 8114A Service Manual page 141

Programmable pulse generator 100 v/2 a
Table of Contents

Advertisement

- 1 4 8
Character data not allowed
HP-IB message = "No MINiMAX allowed while
checks are off"
Local message = "No MIN/MAX while checks are
off"
Description:
An attempt was made to set a parameter to MINIMAX
checks were switched off.
This does not work because other parameters on
which the MIN/MAX result depends might already be
outside no-warn range.
- 2 0 0
Execution error
HP-IB message = "No signal at CLK-IN could be
detected for frequency measurement"
Local message = "No signal detected at CLK-INput"
Description:
An attempt was made to measure the external
clock frequency but no signal was detectable on the
clock-in connector.
- 2 0 0
Execution error
HP-IB message = "Frequency at CLK-INput too
low. "
Local message = "Frequency at CLK-INput too
low. "
Description:
An attempt was made to measure the external clock
frequency but the detected signal frequency was too
low to allow an accurate measurement. (Currently
the limit is 1000 Hz.)
- 2 0 0
Execution error
HP-IB message = "Frequency at CLK-INput too
high. "
Local message = "Frequency at CLK-INput too
high. "
Description:
An attempt was made to measure the external clock
frequency but the detected signal frequency was too
high to allow an accurate measurement.
HP 8 114A Service Guide
Troubleshooting