Test 2: Width; Test Specifications; Equipment Needed; Procedure - HP 8114A Service Manual

Programmable pulse generator 100 v/2 a
Table of Contents

Advertisement

Test 2: Width

Test Specifications

Equipment Needed

Procedure

10.0 ns to 150 ms
3 digits, best case 100 ps
Accuracy
0.03% +25 ps (0.05% +25 ps for 50 ns <width ~100
Digitizing Oscilloscope with Accessories
Counter
Cable, 50 0, coaxial, BNC
Connecting FIP 8114A to the Scope
. C:ijf.jTI t.JfJ[ll_i::; F'!ji_S&
n :jj.r,gj.t+:
3. On the HP 8114A press m and set up UJTPUT page as shown
in the following illustration:
4. Set the Digitizing Oscilloscope HP 54121T:
Press [AUTOSCALE)
n
Select the Display menu and set the Number of Averages to 32
n
Select the delta V menu and turn the voltage markers On
n
Set
the preset levels to 50% -50% and press
n
A D A P T O R
Configuring the Output Page
(
]
AUTO LEVEL SET
Testing the HP g114A 3-7