Test 5.2: Width Jitter
Test Specifications
0.03% + 25 ps (0.05% + 25 ps for 50 ns < width <
100 ns)
Equipment Needed
Digitizing Oscilloscope with Accessories
Delay Line (22 ns)
Power Splitter
All cables 50 0, coaxial, BNC, 122cm (4ft)
High Power Attenuator
Procedure
1. Connect HP 8114A to the Scope as shown:
Equipment Set-up for Jitter Test
Testing the HP g114A