4. Take the DVM readings for the following baseline settings (VB).
Multiply the reading by the attenuation factor G1 [derived from the
amplitude test, test 6).
B = VBxG,
Compare the calculated value with the given limits:
E laseline
- 2 5 v
- 2 0 v
-10 v
- 5 v
- 2 v
l o v
+ 10 v
ts
Li
m i
m i n i m u m
m a x i m u m
-25.355 V
-24.645 V
-20.305 V
-19.696 V
7 - 2
-10.205 v
-9.795 V
7 - 3
-6.155 v
- 4 . 4 3 5 v
7 - 4
-2.125 V
-1.876 V
- 0 . 1 0 5 v
7 - 6
7 - 7
7 - K
+ 4.845 V
7 - 9
7 10
Testing the HP 8114A