50 0, 10 W Feedthrough Termination; Getting Started; Instrument Serial Numbers - HP 8114A Service Manual

Programmable pulse generator 100 v/2 a
Table of Contents

Advertisement

50 R, 10 W Feedthrough Termination

Getting Started

Instrument Serial Numbers

The following figure provides a schematic and a parts list except for
the case. The case must provide shielding and maintain grounding
integrity.
B N C
50 Ohm, 0.1% , 10 W Feedthrough Termination
The following parts are required:
1. RI = 53.60, l%, 10 W; HP Part Number: 0699-0146.
2. R2 = 200 0, lo%, 0.5 W, Variable trimmer; HP Part Number:
2 100-3350.
3.
R3 = 681 0, l%, 0.5 W; HP Part Number: 0757-0816.
4. BNC (M): HP Part Number: 1250-0045.
5. BNC (F): HP Part Number: 1250-0083.
You will need to write the serial numbers of the instrument. These
can
be found as follows:
Press m, SERIAL #
The HP 8114A display lists the instrument's product and serial
numbers.
The display on your instrument should look similar to this:
Figure 3- 1.
Testing the HP 8114A

Hide quick links:

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents