HP 8114A Service Manual page 153

Programmable pulse generator 100 v/2 a
Table of Contents

Advertisement

File name error
-257
Local message = "Permission denied"
Description:
An attempt was made to do one of the following:
delete a directory
n
change directory to a file.
Media protected
-258
HP-IB message = "Drive A: Memory Card write
protected"
Local message = "Drive A: Memory Card write
protected"
Battery low
-258
HP-IB message = "Drive A: Memory Card battery is
low"
Local message = "Drive A: Memory Card battery is
low"
Description:
The battery of the Memory Card should be replaced.
Device specific error
-300
HP-IB message = "&OVERVOLZ4GIiX Output
disabled"
Local message = "&OVERVOLXAGE& Output
disabled"
Description:
The output was disabled by the overvoltage
protection circuitry automatically.
Device specific error
-300
HP-IB message = "Output Bd.: Can't reset the
interrupt circuit"
Local message = "Output Bd.: Interrupt reset
failed"
Description:
The interrupt line can't, be cleared by the software.
There must be a problem on the output board.
HP 8114A Service Guide
Troubleshooting 5-35