HP 8114A Service Manual page 158

Programmable pulse generator 100 v/2 a
Table of Contents

Advertisement

- 3 3 0
HP-IB message = "Internal Serial Device Bus
failed"
Local message = "Internal Serial Device Bus failed"
Description:
Internal serial device bus traffic over feedback path
has failed.
- 3 3 0
HP-IB message = "[device-dependent info]"
Local message = "[device-dependent, info"
- 3 3 0
HP-IB message = "SELFTEST
Local message = "SELFTEST
- 3 3 0
HP-IB message = "SELFTEST: INSTALLED
BOARDS OK"
Local message = "SELFTEST: INSTALLED BOARDS
OK"
- 3 3 0
HP-IB message = "SELFTEST FAILED"
Local message = "SELFTEST FAILED"
- 3 3 0
HP-IB message = "Failure in PERIOD circuitry"
Local message = "Failure in PERIOD circuitry"
Description:
The Period TIGER IC may be defective.
- 3 3 0
HP-IB message = "Failure in DELAY circuitry"
Local message = "Failure in DELAY circuitry"
Description:
The Delay TIGER IC may be defective.
- 3 3 0
HP-IB message = "Failure in WIDTH circuitry"
Local message = "Failure in WIDTH circuitry"
Description:
The Width TIGER IC may be defective.
P-BOARD OK"
U
P-BOARD OK"
U

Hide quick links:

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents