6.

WHEN YOU FINISH USING THE METER

Remove its' probe from the flue - THE PROBE WILL BE HOT - and let it cool. Do
not put the probe in water which will be sucked into the meter, damaging its' pump
and sensors.
When the meter's readings return to ambient levels, switch it off. The meter counts
down from 30 before switch off with the pump running to self clean its sensors.
7.

METER PROBLEM SOLVING

If any problems are not solved with these solutions, contact us or an authorized repair
center.
Fault symptom
• Oxygen too high
• CO 2 too low
• Oxygen reading (- - - -)
• CO reading (- - - -)
• Batteries not holding charge
• Meter not running on mains
adapter.
• Meter does not respond to flue
gas
• Net temperature or Efficiency
calculation incorrect.
• Flue temperature readings erratic
• T flue or T nett displays (- - - -)
• X-Air, EFF, COa or CO2 display
(- - - -)
• Meter just continually beeps
12
Causes / Solutions
• Air leaking into probe, tubing, water trap, connectors
or internal to meter.
• Oxygen cell needs replacing.
• Meter was stored in a cold environment and is not at
normal working temperature.
• Oxygen cell or CO sensor needs replacing.
• Pump is switched off
• Batteries exhausted.
• AC charger not giving correct output.
• Fuse blown in charger plug.
• Particle filter blocked.
• Probe or tubing blocked.
• Pump not working or damaged with contaminants.
• Ambient temperature set wrong during Automatic
Calibration.
• Temperature plug reversed in socket.
• Faulty connection or break in cable or plug.
• Probe not connected.
• Oxygen reading is above 18%
• Turn dial back to MENU and press ENTER
KANE200 series manual