Automatic Recording
Press Mode key on the screen or the control panel to toggle between Auto 1+1, Auto 3, Auto 3+1
and Auto 6.
Press Start/Stop key on the control panel to start recording for 3-12 seconds (Refer to section 2 "User
Setting Menu").
It is not possible to toggle between different leads using Lead► and ◄Lead keys.
There are four modes in which only specific leads can be recorded first:
Auto 1+1: lead I
Auto 3 and Auto 3+1: leads I, II, III
Auto 6: I, II, III, aVF, aVR, aVL
After ending the Time of Recording of lead/leads, the record will continue for the next lesd/lead group
automatically.
By pressing the Start/Stop key at any time, record will stop.
■ Auto 1+1: Select this mode and press the Start/Stop key to start recording.
In the recording paper the first waveform indicates the waveform of selected lead and the second one
is the waveform of Rhythm lead.
■ Auto 3: Select this mode and press the Start/Stop key to start recording.
■ Auto 3+1: Select this mode and press the Start/Stop key to start recording.
In the recording paper the first three waveforms are the waveforms of selected leads and the last one
is the waveform of Rhythm lead. The recording duration will be according to the "Rec Time".
■ Auto 6: Select this mode and press the Start/Stop key to start recording.
45