Connecting The Scale - Sartorius EVO1Y Operating Instructions Manual

Paint-mixing scales for use in potentially explosive atmospheres of zone 2
Table of Contents

Advertisement

4.4.2

Connecting the Scale

Connecting the Scale
Insert the USB cable plug into the USB socket on the back of the display.
t
Loosen the screw (1) on the lock.
t
Swivel the lock over the USB cable.
1
t
Re-tighten the lock screw.
t
Laying the USB Cable
Lay the USB cable through the cable holders on the back of the scale.
t
Connecting the Grounding Cable
This explosion-protected system should be set up according to commonly accepted
technical standards. The applicable national electrical code and safety regulations for
your particular country must be observed.
Before commissioning the scale, a check must be carried out by or under the
supervision of a qualified electrician to ensure that the system is in good working
order.
Check whether or not the competent authorities (e.g., industrial supervisory board)
need to be informed. It is also necessary to carry out inspections of the system
during operation.
Inspection intervals should be such that any significant defects that may occur can
be identified in good time. Inspections should be carried out at least once every
three years. The applicable requirements and guidelines should also be observed
during operation.
Establish a low resistance connection from the scale to a customer-supplied
equipotential bonding conductor connection via the equipotential bonding
conductor connection on the device using a suitable grounding cable with a
gauge of at least 4 mm
(not included).
2
EVO1Y | LAB1Y Operating Instructions
Installation
19