Operation; Switching The Device On/Off; Locking The Keypad/Weight Display; Zeroing/Taring - Sartorius EVO1Y Operating Instructions Manual

Paint-mixing scales for use in potentially explosive atmospheres of zone 2
Table of Contents

Advertisement

Operation

0
30
22
EVO1Y | LAB1Y Operating Instructions
6
Operation
6.1

Switching the Device On/Off

Switching On
Briefly press the j key (on/standby).
t
The automatic self-test runs. This ends when the display shows 0.0 g.
y
If another value is displayed: Set the scale to zero via the 0 key.
t
Switching Off
Press and hold the j key for several seconds (on/standby).
t
The scale switches to standby mode.
y
6.2

Locking the Keypad/Weight Display

Locking
Briefly press the j key (on/standby) to lock the keypad and turn off the weight
t
display.
The padlock symbol H flashes for several seconds.
y
Press the padlock symbol H while flashing to activate the lock.
t
The keypad/weight display is locked and the padlock symbol lights up continually.
y
Unlocking
Press the padlock symbol H to deactivate the lock.
t
Enter the password (if set) (see Chapter 6.7.2.6, page 29).
t
The lock is deactivated.
y
6.3

Zeroing/Taring

Briefly press the 0 key.
t
6.4

Adjustment

The scale needs to be re-leveled and then adjusted each time its setup location is
changed.
Warm-up Time
To ensure accurate results are delivered, the scale must warm up for at least
30 minutes after initial connection to the power supply.
Only after this time will the scale have reached the required operating temperature.
Wait approx. 30 minutes after connecting the scale to the power supply before
t
adjusting the scale and weighing.
Using a Scale Verified for Use in Legal Metrology:
− Ensure that there is a warm-up time of at least 24 hours after initial connection
to the power supply.