Ground (Earth); Connecting The Scale To Ac Power - Sartorius Signum: Option Y2 SIWR Installation Instructions And Safety Information

Table of Contents

Advertisement

6 6
Grounding
For models with analog platform (option A15) and
internal battery pack (option L9) only
§ Ground the weighing platform.
Securely connect the cable to the grounding terminal.
Assembly:
– Toothed lock washer, cable, toothed lock washer
– Screw, or nut with locknut
§ Additionally, secure the cable with a clip
(strain relief).
For SIW..DCP models:
§ Ground the weighing platform with the column.
Securely connect the cable to the grounding terminal.
Assembly:
– Toothed lock washer, cable, toothed lock washer
– Screw, or nut with locknut
§ Additionally, secure the cable with a clip
(strain relief).
For SIWS..BBP models:
§ Ground the indicator.
Connect the cable to the grounding terminal.
Assembly:
– Toothed lock washer, cable, toothed lock washer
– Nut
– Locknut
Power Connection
!
Make sure that the voltage rating printed on this
unit matches the voltage at the place of installation.
If the voltage specified on the label does not match
the rating or standard you use, please contact your
local Sartorius office or dealer.
!
Make sure to observe the safety instructions.
§ Connect the equipment to AC power.