11
Lay cable and insert front panel
Pull the front panel
Plug the flat control cable
Lay the motor cable
bottom of the housing.
Note
NOTE ON LAYING THE MOTOR CABLE
It is much easier to lay the cables if the lamp holder is temporarily
dismantled / removed from the housing.
At the narrow points (see red dotted ellipse in Fig. 23 on page 26) , take
special care to avoid damaging the electronic components or cables.
Plug the motor cable
Insert the front panel
Ensure that the flat control cable is laid so that it runs through the two
recesses on the underside of the sheet metal
(see the two red dashed circles in in Fig. 23 on page 26).
X702*
X704
X800
3.1
4.1
Fig. 22
Lay the cable and insert the front panel
slightly out of the housing.
4
into slot X704
4.2
so that it runs below the lamp cable along the
3.2
into slot X800
.
3.2
3.1
very carefully into the housing from above
2
2.1
4.2
3.2
Changing the motor
on the "FBG PowerBoard".
4.1
1
of the front panel
2
1
2.1
2
.
25