Zeiss AURA handheld NIR Maintenance And Repair Manual page 16

Advertisement

Changing the lamp cable (optional)
Requirements
Procedure
16
5
Changing the lamp cable (optional)
The following steps should only be performed if required.
Scheduled regular replacement of the lamp cable is not required.
The display has been removed (see section "3 Changing the display" on page 8)
The front plate
has been removed
1
(see section "6 Changing the motor", steps 1 to 2)
The lamp holder
2
(see section "4 Replacing the halogen lamp")
Put the following spare parts ready:
– lamp cable, order no. 000000-2224-159
1
Disconnect the lamp cable from the halogen lamp
Disconnect the lamp cable from the halogen lamp
At the narrow points (see red dotted ellipse in the figure below), take special
care to avoid damaging the electronic components (or the new lamp cable
during installation).
1
Fig. 10
Disconnect the lamp cable from the halogen lamp
Maintenance and Repair Manual | AURA
AURA
has been disassembled
2
3
4
handheld NIR - Rev. 2 | EN_40_070_0030I
®