INSTALLATION (continued) 
5. Stand  Pipe:    A  standpipe  maybe  used  to  dampen  turbulence  or  when  foam  is  present  in  the 
application.   
a. Pipe can be made of any material. 
b. Select a minimum 3" ID pipe for the stand pipe. 
i. A 2" pipe is usable with the LVU809 series, but is the minimum. 
ii. Pipes larger than 3" can also be used. 
c. Use a coupling and reducer bushing to attach the LVU800 Series to the pipe. 
i. With the LVU809 series, be sure to use a plastic reducing bushing such as the 2" Thread 
ii. With  the  LVU816,  LVU826  or  LVU832  series,  be  sure  to  use  a  plastic  reducing  bushing 
d. The pipe length should run the measurement span and the bottom of the pipe should remain 
submerged at all times to prevent foam from entering the pipe. 
e. Cut a 45°notch at the bottom of the pipe and drill a 1/4"pressure equalization hole in the dead 
band.   
f. The pumps should not drive liquid past the open end of the stand pipe which causes the liquid 
in the pipe to oscillate. 
 
 
 
x 1" Thread fitting (ex: LVU800‐2N80) or 2" Slip x 1" Thread fitting (ex: LVU800‐2S80).  
such as the 3" Thread x 2" Thread fitting (ex: LVU800‐3N80) or 3" Slip x 2" Thread fitting 
(ex: LVU800‐3S80).  
 
 
19 
 