GETTING STARTED (continued) 
How to set a Reverse the Current Output: 
 
How to set the SAFE setting: 
 
 
 
The default for LVU800 Series is to have 4 mA at Empty and 20 mA at 
Full.  This is the normal (NORM) setting.  The output can be reversed 
(REV) with 20 mA at Empty and 4 mA at Full.   
1. In MENU mode, select the TANK item. 
 
2. When REV MA appears, press SELECT. 
3. When REV appears, press SELECT. 
4. When EXIT appears, press SELECT return to Top Level Menu. 
LVU800  Series  is  now  in  the  Reverse  mode.    To  switch  back  to  the 
Normal mode, follow the instructions above and select NORM under 
step 3.  
The default for Fail‐Safety (LOST) can be preset.  The choices are 4 mA, 
20 mA, 21 mA, 22 mA and HOLD.     
 
1. In MENU mode, select the SAFE item. 
2. When the required setting appears, press SELECT. 
3. When EXIT appears, press SELECT return to Top Level Menu. 
 